Gas Flow Problems; Table 11-2: Relay Board Status Leds - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
LED
Color
D1
red
Watchdog Circuit; I
operation.
D2
yellow
Relay 0 - sample chamber
heater
D3
yellow
Relay 1 – SO
heater
1
D4
yellow
Relay 2 - manifold heater
D5
yellow
Relay 3 - IZS heater
D6
yellow
Relay 4 - Spare
2
D7
green
Valve 0 - zero/span valve
status
2
D8
green
Valve 1 - sample/cal valve
status
D9
green
Valve 2 - auto-zero valve
status
D10
green
Valve 3 - SO/SO
status
D11
green
Valve 4 - Spare
D12
green
Valve 5 - Spare
D13
green
Valve 6 - Spare
D14
green
Valve 7 - Spare
D15
green
Mosfet1-Unused
D16
Green
Mosfet2-Unused
1
Special configurations only
2
Only active for instruments with Z/S valve or IZS options installed

11.2. Gas Flow Problems

The standard M100E has one main flow path. With the IZS option installed, there is a
second flow path flow path through the IZS oven that runs whenever the IZS is on
standby to purge SO
2
available from the front panel. The full flow diagrams of the standard configuration
(Figure 3-10) and with options installed (Figure 5-2 and 5-3) help in trouble-shooting
flow problems. In general, flow problems can be divided into three categories:
Flow is too high
Flow is greater than zero, but is too low, and/or unstable
Flow is zero (no flow)
045150102 Rev XB1

Table 11-2: Relay Board Status LEDs

FUNCTION
STATUS
2
C bus
Continuously
ON or OFF
Continuously
ON or OFF
converter
Continuously
2
ON or OFF
Continuously
ON or OFF
Continuously
ON or OFF
N/A
Continuously
ON or OFF
Continuously
ON or OFF
Continuously
ON or OFF
valve
Continuously
x
ON or OFF
N/A
N/A
N/A
N/A
N/A
N/A
from the oven chamber. The IZS flow is not measured and is not
TROUBLESHOOTING & REPAIR
FAULT
INDICATED FAILURE(S)
Failed or halted CPU; faulty motherboard,
keyboard, relay board; wiring between
motherboard, keyboard or relay board; +5
V power supply.
Heater broken, thermistor broken
Heater broken, thermocouple broken
Heater broken, thermistor broken
Heater broken, thermistor broken
N/A
Valve broken or stuck, valve driver chip
broken
Valve broken or stuck, valve driver chip
broken
Valve broken or stuck, valve driver chip
broken
Valve broken or stuck, valve driver chip
broken
N/A
N/A
N/A
N/A
N/A
N/A
203