Operating Instructions
To access and navigate the VARS menu, use the following key sequence.
SAMPLE*
< TST TST > CAL
SETUP X.X
CFG DAS RNGE PASS CLK
SETUP X.X
COMM
VARS
SETUP X.X
8
1
SETUP X.X
SETUP X.X
PREV
SETUP X.X
PREV
SETUP X.X
PREV
SETUP X.X
PREV
SETUP X.X
PREV
SETUP X.X
PREV
SETUP X.X
PREV
72
RANGE = 500.000 PPB
SO2 =X.XXX
PRIMARY SETUP MENU
MORE
SECONDARY SETUP MENU
DIAG
ENTER VARS PASS: 818
8
ENTR
DAS_HOLD_OFF=15.0 Minutes
0 )
JUMP
PRNT EXIT
NEXT
EDIT
1 ) TPC_ENABLE=ON
JUMP
PRNT EXIT
NEXT
EDIT
2)RCELL_SET=50.0 DegC
JUMP
PRNT EXIT
NEXT
EDIT
3)
IZS_SET=50.0 DegC
JUMP
EDIT
PRNT EXIT
NEXT
4 ) DYN_ZERO=ON
JUMP
PRNT EXIT
NEXT
EDIT
5) DYN_SPAN=ON
JUMP
PRNT EXIT
NEXT
EDIT
6) CONC_PRECUISION : 3
JUMP
PRNT EXIT
NEXT
EDIT
7) CLOCK_ADJ=0 Sec/Day
JUMP
PRNT EXIT
NEXT
EDIT
SETUP
EXIT
EXIT
EXIT ignores the new setting.
ENTR accepts the new setting.
EXIT
SETUP X.X
1
5
.0
SETUP X.X
ON
SETUP X.X
ON
SETUP X.X
ON
SETUP X.X
AUTO
0
SETUP X.X
CLOCK_ADJ=0 Sec/Day
+
0
0
Model 100E Instruction Manual
DAS_HOLD_OFF=15.0 Minutes
ENTR EXIT
Toggle this keys to change setting
TPC_ENABLE=ON
ENTR EXIT
Toggle this keys to change setting
DO NOT change
theses set-points
unless
specifically
instructed to by
T-API Customer
Service.
DYN_ZERO=ON
ENTR EXIT
Toggle this keys to change setting
DYN_SPAN=ON
ENTR EXIT
Toggle this keys to change setting
CONC_PRECUISION : 3
1
2
3
4
ENTR EXIT
Toggle these keys to change setting
ENTR EXIT
Toggle tese keys to change setting
045150102 Rev XB1