Table 3-1: Analog Output Connections Pin Function - Teledyne 2000RS Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Installation
Analog Outputs: There are four DC output signal pins—two pins
per output. For polarity, see Table 3-1. The outputs are user selectable
(either 0-1 vdc or 4-20 ma) but must be selected from the system menu:
 0–1 VDC % of Range: Voltage rises linearly with increasing
 0–1 VDC Range ID:
 4–20 mA DC % Range: Current increases linearly with
 4–20 mA DC Range ID: 8 mA = Low Range, 12 mA = Medium

Table 3-1: Analog Output Connections Pin Function

Pin
3
+ Range ID, 4-20 mA, floating
4
– Range ID, 4-20 mA, floating
5
+ % Range, 4-20 mA, floating
6
– % Range, 4-20 mA, floating
8
+ Range ID, 0-1 VDC
23
– Range ID, 0-1 VDC, negative ground
24
+ % Range, 0-1 VDC
7
– % Range, 0-1 VDC, negative ground
Alarm Relays: The nine alarm-circuit connector pins connect to
the internal alarm relay contacts. Each set of three pins provides one set
of Form C relay contacts. Each relay has both normally open and
normally closed contact connections. The contact connections are shown
in Table 3-2. The contacts are capable of switching up to 1 amp at 250 V
ac into a resistive
load.
Teledyne Analytical Instruments
concentration, from 0 V at 0 ppm to 1 V
at full scale ppm. (Full scale = 100% of
programmable range.)
0.25 V = Low Range, 0.5 V = Medium
Range, 0.75 V = High Range, 1 V = Air
Cal Range.
increasing concentration, from 4 mA at
0 ppm to 20 mA at full scale ppm. (Full
scale = 100% of programmable range.)
Range, 16 mA = High Range, 20 mA =
Air Cal Range.
Function
The connectors are:
2000RS
20