3.1.1.1. Connecting the Analog Outputs

Attach a strip chart recorder and/or data-logger to the appropriate contacts of the analog
output connecter on the rear panel of the analyzer.
The A1 and A2 channels output a signal that is proportional to the SO
the sample gas.
The output, labeled A4 is special. It can be set by the user (Section 6.7.10) to output
any one of the parameters accessible through the <TST TST> keys of the units sample
display.
Pin-outs for the Analog Output connector at the rear panel of the instrument are:
PIN
1
2
3
4
5
6
7
8
The default analog output voltage setting of the M100E UV Fluorescence SO2 Analyzer is
0 – 5 VDC with a range of 0 – 500 ppb.
TO change these settings, see Sections 6.5
An optional Current Loop output is available for each (See Section 5.2).
28
ANALOG OUT
A1
A2
+
-
+
-
Figure 3-3:
Analog Output Connector
Table 3–1:
Analog output Pin Outs
ANALOG OUTPUT
A1
A2
A3
A4
Model 100E Instruction Manual
A3
A4
+
-
+
-
STANDARD
CURRENT LOOP
VOLTAGE
OUTPUT
V Out
Ground
V Out
Ground
Not Available
Not Available
Not Available
Not Available
V Out
Ground
concentration of
2
OPTION
I Out +
I Out -
I Out +
I Out -
I Out +
I Out -