Table A-3: M100E Test Functions, Revision B.3 - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
TEST FUNCTION
RANGE
RANGE=500.0 PPB
RANGE1
RANGE1=500.0 PPB
RANGE2
RANGE2=500.0 PPB
STABILITY
STABIL=0.0 PPB
2
RESPONSE
RSP=1.11(0.00) SEC
5
VACUUM
VAC=9.1 IN-HG-A
SAMPPRESS
PRES=29.9 IN-HG-A
SAMPFLOW
SAMP FL=700 CC/M
PMTDET
PMT=762.5 MV
NORMPMTDET
NORM PMT=742.9 MV
UVDET
UV LAMP=3457.6 MV
LAMPRATIO
LAMP RATIO=100.0 %
STRAYLIGHT
STR. LGT=0.1 PPB
DARKPMT
DRK PMT=19.6 MV
DARKLAMP
DRK LMP=42.4 MV
SLOPE
SLOPE=1.061
OFFSET
OFFSET=250.0 MV
HVPS
HVPS=650 VOLTS
RCELLDUTY
RCELL ON=0.00 SEC
RCELLTEMP
RCELL TEMP=52.1 C
BOXTEMP
BOX TEMP=35.5 C
PMTTEMP
PMT TEMP=7.0 C
IZSDUTY
IZS ON=0.00 SEC
IZSTEMP
IZS TEMP=52.2 C
SO2
SO2=261.4 PPB
TESTCHAN
TEST=3721.1 MV
CLOCKTIME
TIME=10:38:27
1
The name is used to request a message via the RS-232 interface, as in "T BOXTEMP".
2
Engineering software.
3
Current instrument units.
5
M100EH
045150110 Rev B.3
Table A-3:
M100E Test Functions, Revision B.3
MESSAGE TEXT
3
3
3
3
APPENDIX A-3: Warnings and Test Functions,
DESCRIPTION
D/A range in single or auto-range modes.
D/A #1 range in independent range mode.
D/A #2 range in independent range mode.
Concentration stability (standard deviation based on
setting of STABIL_FREQ and STABIL_SAMPLES).
Instrument response. Length of each signal processing
loop. Time in parenthesis is standard deviation.
Vacuum pressure.
Sample pressure.
Sample flow rate.
Raw PMT reading.
PMT reading normalized for temperature, pressure,
auto-zero offset, but not range.
UV lamp reading.
UV lamp ratio of current reading divided by calibrated
reading.
Stray light offset.
PMT dark offset.
UV lamp dark offset.
Slope for current range, computed during zero/span
calibration.
Offset for current range, computed during zero/span
calibration.
High voltage power supply output.
Sample chamber temperature control duty cycle.
Sample chamber temperature.
Internal chassis temperature.
PMT temperature.
IZS temperature control duty cycle.
IZS temperature.
SO
concentration for current range.
2
Value output to TEST_OUTPUT analog output, selected
with TEST_CHAN_ID variable.
Current instrument time of day clock.
Revision B.3
241