Figure 5-3: Pneumatic Diagram Of The M100E With Izs Options Installed - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Optional Hardware and Software
INSTRUMENT CHASSIS
SAMPLE GAS
INLET
EXHAUST GAS
OUTLET
SPAN GAS INLET
ZERO AIR INLET
ZERO/SPAN
VALVE
2
3
1
IZS
Permeation Tube
NO
Source
2
Figure 5-3:
The state of the IZS valves can also be controlled:
Manually from the analyzer's front panel by using the SIGNAL I/O controls
located under the DIAG Menu (Section 6.7.2),
By activating the instrument's AutoCal feature (Section 7.8),
Remotely by using the external digital control inputs (Section 6.8.2 and Section
7.7.1), or
Remotely through the RS-232/485 serial I/O ports (see Appendix A-6 for the
appropriate commands).
External Zero Air Scrubber
The IZS option includes an external zero air scrubber assembly that removes all SO
zero air source. The scrubber is filled with activated charcoal.
The Permeation Source
50
EXHAUST TO OUTER
LAYER OF KICKER
FLOW
CONTROL
ASSY
SAMPLE/CAL
VALVE
3
2
1
FLOW
SENSOR
Pneumatic Diagram of the M100E with IZS Options Installed.
Model 100E Instruction Manual
SAMPLE
CHAMBER
PMT
SAMPLE
PRESSURE
SENSOR
FLOW / PRESSURE
SENSOR PCA
PUMP
UV
LAMP
HYDROCARBON
SCRUBBER
(KICKER)
SAMPLE
FILTER
the
2
045150102 Rev XB1