Figure 11-4: Pre-Amplifier Board Layout - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
Turn the gain adjustment potentiometer 12 turns clockwise to its maximum
setting.
While feeding 400 ppb SO
the STABIL value is below 0.5 ppb, look at the front panel and scroll to the NORM
PMT value. This value should always be two times the span gas concentration in
ppb. With 400 ppb SO
calibrated analyzer.
Set the HVPS coarse adjustment to its minimum setting (0). Set the HVPS fine
adjustment switch to its maximum setting (F).
Set the HVPS coarse adjustment switch to the lowest setting that will give you
more than 800 mV NORM PMT signal. The coarse adjustment typically increments
the NORM PMT signal in 100-300 mV steps.
Adjust the HVPS fine adjustment such that the NORM PMT value is just above 800
mV. It may be necessary to go back and forth between coarse and fine
adjustments if the proper value is at the threshold of the min/max coarse setting.
Do not overload the PMT by accidentally setting both adjustment switches to
their maximum setting. This can cause permanent damage to the PMT.
Adjust the NORM PMT value with the gain potentiometer down to 800±10 mV. This
is the final very-fine adjustment.
045150102 Rev XB1

Figure 11-4: Pre-Amplifier Board Layout

(or 80% range value) to the analyzer and waiting until
2
, the NORM PMT should show 800 mV on a properly
2
NOTE
TROUBLESHOOTING & REPAIR