Model 100E Instruction Manual
Setup Variable
3
VAC_SAMP_RATIO
SAMP_PRESS_SET
SAMP_PRESS_SLOPE
3
VAC_PRESS_SET
BOX_SET
PMT_SET
045150110 Rev B.3
Numeric
Default Value
Units
—
0.53
"Hg-A
29.92
Warnings:
15 to 35
—
1
"Hg
6
Warnings:
3 to 10
ºC
30
Warnings:
8 to 50
ºC
7
Warnings:
2 to 12
APPENDIX A-2: Setup Variables For Serial
Value Range
0.1 to 2
0 to 100
0.5 to 1.5
0 to 100
5 to 60
0 to 40
I/O, Revision B.3
Description
Maximum vacuum
pressure / sample
pressure ratio for valid
sample flow calculation.
Sample pressure set point
for pressure compensation
and warning limits.
Sample pressure slope
correction factor (adjusted
pressure = measured
pressure x slope).
Vacuum pressure set point
for pressure compensation
and warning limits.
Box temperature warning
limits. Set point is not
used.
PMT temperature set point
and warning limits.
233