Teledyne 100E Instruction Manual page 132

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Calibration Procedures
STEP TWO: Set the expected
instrument is set for single (SNGL) range mode with a reporting range span of 500 ppb.
The SO
span concentration
2
values automatically default
to 450.0 Conc.
To change this value to the
actual concentration of the
span gas, enter the number
by pressing the key under
each digit until the expected
value appears.
The span gas concentration
should always be 90% of the
selected reporting range
EXAMPLE
Reporting range = 800 ppb
Span gas conc.= 720 ppb
130
SO
span gas concentrations. In this example the
2
SAMPLE*
RANGE = 500.000 PPB
< TST TST >
CAL
M-P CAL
RANGE = 500.000 PPB
< TST TST >
ZERO
CONC
M-P CAL
SO2 SPAN CONC: 450.0 Conc
0
0
0
4
5
Model 100E Instruction Manual
SO2 =X.XXX
SETUP
SO2 =X.XXX
EXIT
.0
ENTR EXIT
This sequence causes the
analyzer to prompt for the
expected NO and NO
x
span concentration.
EXIT ignores the new setting
and returns to the previous
display.
ENTR accepts the new setting
and returns to the
previous display..
045150102 Rev XB1