Calibration Procedures
7.6. Manual Calibration Checks with IZS or Zero/Span
Valves
Zero and span checks using the zero/span valve or IZS option are similar to that
described in Section 7.3, except:
•
On units with an IZS option installed, zero air and span gas are supplied to the
analyzer through the zero gas inlet and from ambient air.
•
On units with a zero/span valve option installed, zero air and span gas are
supplied to the analyzer through the zero gas and span gas inlets from two
different sources.
•
The zero and calibration operations are initiated directly and independently with
dedicated keys CALZ and CALS.
To perform a manual calibration check of an analyzer with a zero/span valve or IZS
Option installed, use the following method
STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.
MODEL 700
Gas Dilution Calibrator
(with O
generator option)
3
MODEL 701
Calibrated
SO
gas
2
(At high
concentration)
Figure 7-3:
136
Source of
SAMPLE Gas
VENT
External Zero
Air Scrubber
Zero Air
Generator
Source of
SAMPLE Gas
Ambient
Air
Setup for Manual Calibration Check with Z/S Valve or IZS Option
Model 100E Instruction Manual
Zero/Span Valves – Option 50
VENT if input is pressurized
Sample
Exhaust
Span
Zero Air
Filter
Internal Zero/Span Option (IZS) – Option 51
VENT if input is pressurized
Sample
MODEL
Exhaust
100E
Span
Zero Air
MODEL
100E
045150102 Rev XB1