7.2. Manual Calibration

The following section describes the basic method for manually calibrating the Model 100E
SO
analyzer.
2
ENTR
Pressing the
for OFFSET and SLOPE (instrument response curve) and alters the instrument's
If you wish to perform a calibration CHECK, do not press ENTR and see Section 7.3.
STEP ONE: Connect the sources of zero air and span gas as shown below.
Calibrated
SO
GAS
2
(At high
concentration)
MODEL 701
Zero Air
Generator
Calibrated
SO
GAS
2
(At span gas
concentration)
MODEL 701
Zero Air
Generator
Figure 7-1:
045150102 Rev XB1
Calibration and Calibration Check
key during the following procedure re-calculates the stored values
calibration.
MODEL 700 Gas
Dilution
Calibrator
(with Ozone Bench
VENT
Needle valve to
control flow
Valve
Setup for Manual Calibration without Z/S valve or IZS Option
No Valve Options Installed
Option)
Sample
Exhaust
Span
Zero Air
OR
Sample
Exhaust
VENT
Span
Zero Air
Calibration Procedures
Source of
SAMPLE Gas
Removed
during
Calibration
MODEL
100E
Source of
SAMPLE Gas
Removed
during
calibration
MODEL
100E