Table 3-1: Analog Output Connections Pin Function - Teledyne TEXIO PTE6-200 Operating Instructions Manual

Percent oxygen analyzer
Table of Contents

Advertisement

Percent Oxygen Analyzer

Table 3-1: Analog Output Connections Pin Function

Pin
3
+ Range ID, 4-20 mA, floating
– Range ID, 4-20 mA, floating
4
5
+ % Range, 4-20 mA, floating
– % Range, 4-20 mA, floating
6
8
+ Range ID, 0-1 VDC
– Range ID, 0-1 VDC, negative ground
23
24
+ % Range, 0-1 VDC
– % Range, 0-1 VDC, negative ground
7
Alarm Relays: The nine alarm-circuit connector pins connect to
the internal alarm relay contacts. Each set of three pins provides one set
of Form C relay contacts. Each relay has both normally open and
normally closed contact connections. The contact connections are shown
in Table 3-2. Contacts can switch up to 1 amp at 250 V ac into a
resistive load. The connectors are:
Threshold Alarm 1:
• Can be configured as high (actuates when
• Can be configured as failsafe or non-failsafe.
• Can be configured as latching or non-latching.
• Can be configured out (defeated).
Threshold Alarm 2:
• Can be configured as high (actuates when
• Can be configured as failsafe or non-failsafe.
• Can be configured as latching or non-latching.
• Can be configured out (defeated).
Teledyne Analytical Instruments
Function
concentration is above threshold), or low (actuates
when concentration is below threshold).
concentration is above threshold), or low (actuates
when concentration is below threshold).
Installation
25