Name
Message Text
WSYSRES
SYSTEM RESET
WDATAINIT
DATA INITIALIZED
WCONFIGINIT
CONFIG INITIALIZED
WPMT
PMT DET WARNING
WUVLAMP
UV LAMP WARNING
WSAMPFLOW
SAMPLE FLOW WARN
WSAMPPRESS
SAMPLE PRESS WARN
5
WVACPRESS
VACUUM PRESS WARN
WBOXTEMP
BOX TEMP WARNING
WRCELLTEMP
RCELL TEMP WARNING
WIZSTEMP
IZS TEMP WARNING
WPMTTEMP
PMT TEMP WARNING
5
WDARKCAL
DARK CAL WARNING
WHVPS
HVPS WARNING
WDYNZERO
CANNOT DYN ZERO
WDYNSPAN
CANNOT DYN SPAN
WREARBOARD
REAR BOARD NOT DET
WRELAYBOARD
RELAY BOARD WARN
WFRONTPANEL
FRONT PANEL WARN
WANALOGCAL
ANALOG CAL WARNING
240
M100E Warning Messages, Revision B.3
Description
Instrument was power-cycled or the CPU was reset.
Data storage was erased.
Configuration storage was reset to factory configuration
or erased.
PMT detector outside of warning limits specified by
DETECTOR_LIMIT variable.
UV lamp reading outside of warning limits specified by
DETECTOR_LIMIT variable.
Sample flow outside of warning limits specified by
SAMP_FLOW_SET variable.
Sample pressure outside of warning limits specified by
SAMP_PRESS_SET variable.
Vacuum pressure outside of warning limits specified by
VAC_PRESS_SET variable.
Chassis temperature outside of warning limits specified
by BOX_SET variable.
Sample chamber temperature outside of warning limits
specified by RCELL_SET variable.
IZS temperature outside of warning limits specified by
IZS_SET variable.
PMT temperature outside of warning limits specified by
PMT_SET variable.
Dark offset above limit specified by DARK_LIMIT
variable.
High voltage power supply output outside of warning
limits specified by HVPS_SET variable.
Contact closure zero calibration failed while DYN_ZERO
was set to ON.
Contact closure span calibration failed while DYN_SPAN
was set to ON.
Rear board was not detected during power up.
Firmware is unable to communicate with the relay board.
Firmware is unable to communicate with the front panel.
The A/D or at least one D/A channel has not been
calibrated.