Specifications, Approvals and Warranty
2.2. EPA Equivalency Designation
The Model 100E Analyzer is designated as Reference Method Number EQSA-0495-100 as
per 40 CFR Part 53 when operated under the following conditions:
•
Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).
•
Ambient temperature range of 5
•
Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
•
Sample filter: Equipped with PTFE filter element in the internal filter assembly.
•
Sample flow of 650 +/- 65 cc/min.
•
Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 slpm
or better.
•
Software settings:
Dynamic span
Dynamic zero
Dilution factor
AutoCal
Dual range
Auto-range
Temp/Pressure compensation
Under the designation, the analyzer may be operated with or without the following
optional equipment:
•
Rack mount with chassis slides.
•
Rack mount without slides, ears only.
•
Zero/span valve options.
•
Internal zero/span (IZS) option with:
a) SO
2
certified/uncertified.
b) SO
2
certified/uncertified. Under the designation, the IZS option cannot be used
as the source of calibration.
•
4-20mA isolated analog outputs.
•
Status outputs.
•
Control inputs.
•
RS-232 output.
•
Ethernet output.
•
Zero air scrubber.
•
4-20mA, isolated output.
20
permeation tube - 0.4ppm at 0.7 liter per minute;
permeation tube - 0.8 ppm at 0.7 liter per minute;
o
o
C to 40
C.
OFF
OFF
OFF
ON or OFF
ON or OFF
ON or OFF
ON
Model 100E Instruction Manual
045150102 Rev XB1