6.3. Calibration Mode

6.3.1. Calibration Functions

Pressing the CAL key switches the M100E into calibration mode. In this mode, the user
can calibrate the instrument with the use of calibrated zero or span gases.
If the instrument includes either the zero/span valve option or IZS option, the display will
also include CALZ and CALS keys. Pressing either of these keys also puts the instrument
into multipoint calibration mode.
The CALZ key is used to initiate a calibration of the zero point.
The CALS key is used to calibrate the span point of the analyzer. It is
recommended that this span calibration is performed at 90% of full scale of the
analyzer's currently selected reporting range.
Because of their critical importance and complexity, calibration operations are described
in detail in other sections of the manual:
Chapter 7 details basic calibration and calibration check operations.
Chapter 8 describes how to perform an APE protocol calibration.

6.4. Setup Mode

The SETUP mode contains a variety of choices that are used to configure the analyzer's
hardware and software features and to read or configure the internal data acquisition
system (iDAS). For a visual representation of the software menu trees, refer to Appendix
A-1.
Any changes made to a variable during one of the following procedures is not
acknowledged by the instrument until the ENTR Key is pressed
If the EXIT key is pressed before the ENTR key, the analyzer will beep, alerting
the user that the newly entered value has not been accepted.
Some of the functions of the SETUP mode are described in separate sections following
this section. In particular, the RNGE function can be found in Section 6.5, the ACAL
function in Section 7.8,