Teledyne 100E Instruction Manual page 255

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction
Manual
CUSTOMER: _________________________________ PHONE: __________________________________________________
CONTACT NAME: ____________________________ FAX NO. __________________________________________________
SITE ADDRESS: _________________________________________________________________________________________
MODEL 100E SERIAL NO.: ___________________ FIRMWARE REVISION: _______________________________________
1. ARE THERE ANY FAILURE MESSAGES? _______________________________________________________________
________________________________________________________________________________________________________
________________________________________________________________________________________________________
2. PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST
PARAMETERS BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT)
*IF OPTION IS INSTALLED
Parameter
RANGE
STABIL
SAMP PRESS
SAMPLE FLOW
PMT SIGNAL
WITH ZERO AIR
PMT SIGNAL AT
SPAN GAS CONC
NORM PMT AT
SPAN GAS CONC
UV LAMP
LAMP RATIO
STR. LGT
DARK PMT
DARK LAMP
SLOPE
OFFSET
HVPS
RCELL TEMP
BOX TEMP
PMT TEMP
IZS TEMP*
ETEST
OTEST
REF_4096_MV
REF_GND
3. WHAT IS THE SAMPLE FLOW & SAMPLE PRESSURE W/SAMPLE INLET ON REAR OF MACHINE CAPPED?
SAMPLE FLOW -
4. WHAT ARE THE FAILURE SYMPTOMS? ______________________________________________________________
_______________________________________________________________________________________________________
PHONE: (858) 657-9800
045150130 Rev XB1 (04796 Rev B)
Warranty/Repair
Questionnaire
Model 100E
Recorded Value
PPB/PPM
IN-HG-A
3
CM
PPB/PPM
PPB/PPM
Values are in the Signal I/O
CC
SAMPLE PRESS -
TELEDYNE API CUSTOMER SERVICE
Email:
api-customerservice@teledyne.com
TOLL FREE: (800) 324-5190
Acceptable Value
PPB
≤1 PPB WITH ZERO AIR
AMBIENT ± 2 IN-HG-A
/MIN
650 ± 10%
mV
-20 TO 150 mV
mV
0-5000 mV
0-20000 PPB
mV
0-5000 mV
0-20000 PPB
mV
2000 TO 4000 mV
mV
30 TO 120%
PPB
≤ 100 PPB/ ZERO AIR
mV
-50 TO 200 mV
mV
-50 TO 200 mV
1.0 ± 0.3
mV
< 250 mV
V
≈ 400 – 900
ºC
50ºC ± 1
ºC
AMBIENT + ~ 5
ºC
7ºC ± 2º CONSTANT
ºC
50ºC ± 1
mV
2000 mV ± 500
mV
2000 mV ± 1000
mV
4096mv±2mv and Must be Stable
mV
0± 0.5 and Must be Stable
FAX: (858) 657-9816
TELEDYNE
INSTRUMENTS
Advanced Pollution Instrumentation
A Teledyne Technologies Company
IN-HG-A
253