Table 8-1: Activity Matrix For Calibration Equipment & Supplies - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

EPA Protocol Calibration
fully warmed up and its operation has stabilized. During the calibration, the M100E should
be in the CAL mode, and therefore sample the test atmosphere through all components
used during normal ambient sampling and through as much of the ambient air inlet
system as is practicable. If the Instrument will be used on more than one range (i.e.
DUAL or AUTO ranges), it should be calibrated separately on each applicable range.
Calibration documentation should be maintained with each analyzer and also in a central
backup file.
Table 8-1: Activity Matrix for Calibration Equipment & Supplies
Equipment/supplies
Recorder
Sample line and
manifold
Calibration equipment
Working standard SO
cylinder gas or SO
permeation tube
Zero air
Record form
Audit equipment
144
Acceptance limits
Compatible with
output signal of
analyzer; min. chart
width of 150 mm (6
in) is recommended
Constructed of PTFE
or glass
Meets guidelines of
reference 1 and
Section 2.3.2 (Q.A.
Handbook)
Traceable to NIST-
2
SRM meets limits in
2
traceability protocol
for accuracy and
stability (Section
2.0.7, Q.A.
Handbook)
Clean dry ambient
air, free of
contaminants that
cause detectable
response with the
SO
analyzer.
2
Develop standard
forms
Must not be the same
as used for calibration
Model 100E Instruction Manual
Frequency and method
of measurement
Check upon receipt
Check upon receipt
See Section 2.3.9 (Q.A.
Handbook)
Analyzed against NIST-
SRM; see protocol in
Section 2.0.7, Q.A.
Handbook
See Section 2.9.2 (Q.A.
Handbook)
N/A
System must be checked
out against known
standards
Action if requirements
are not met
Return equipment to
supplier
Return equipment to
supplier
Return equipment/
supplies to supplier or
take corrective action
Obtain new working
standard and check for
traceability
Obtain air from another
source or regenerate.
Revise forms as
appropriate
Locate problem and
correct or return to
supplier
045150102 Rev XB1