Teledyne 100E Instruction Manual page 44

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Getting Started
STEP THREE: Perform the zero/span calibration procedure:
The SPAN key now
appears during the
transition from zero to
span.
You may see both keys.
If either the ZERO or
SPAN buttons fail to
appear see Section 11
for troubleshooting tips.
The Model 100E analyzer is now ready for operation.
42
SAMPLE*
RANGE = 500.000 PPB
< TST
>
CAL
TST
SAMPLE
STABIL=X.XXX PPB
< TST TST >
CAL
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
STABIL=X.XXX PPB
< TST TST >
CAL
M-P CAL
STABIL=X.XXX PPB
< TST TST >
CONC
ZERO
M-P CAL
STABIL=X.XXX PPB
< TST TST >
CONC
ENTR
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
M-P CAL
STABIL=X.XXX PPB
< TST TST >
CONC
SPAN
M-P CAL
RANGE = 500.0 PPB
< TST TST >
SPAN CONC
ENTR
M-P CAL
RANGE = 500.0 PPB
< TST TST > ENTR
CONC
Model 100E Instruction Manual
SO2 =X.XXX
Set the Display to show the
This function calculates the
SETUP
stability of the NO/NO
SO2 =X.XXX
SETUP
SO2 =X.XXX
SETUP
SO2 =X.XXX
EXIT
Press ENTR to changes the
SO2 =X.XXX
OFFSET & SLOPE values for the
EXIT
Press EXIT to leave the calibration
unchanged and return to the
SO2 =X.XXX
EXIT
Press ENTR to change the
OFFSET & SLOPE values for the
SO2 =X.XXX
Press EXIT to leave the calibration
EXIT
unchanged and return to the
SO2 =X.XXX
EXIT returns to the main
SAMPLE display
EXIT
045150102 Rev XB1
STABIL test function.
x
measurement
Wait until STABIL
falls below 0.5 ppb.
This may take several
minutes.
SO
measurements.
2
previous menu.
The value of
STABIL may jump
significantly.
Wait until it falls back
below 0.5 ppb.
This may take several
minutes.
SO
measurements.
2
previous menu.