Teledyne 100E Instruction Manual page 236

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

APPENDIX A-2: Setup Variables For Serial I/O,
Revision B.3
Setup Variable
RS232_MODE
BAUD_RATE
MODEM_INIT
RS232_MODE2
234
Numeric
Default Value
Units
BitFlag
0,
5
3
19200
"AT Y0 &D0 &H0
&I0 S0=2 &B0 &N6
&M0 E0 Q1 &W0"
BitFlag
0
Model 100E Instruction Manual
Value Range
Description
0 to 65535
RS-232 COM1 mode flags.
Add values to combine
flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable iChip support
(1)
16 = enable Hessen
protocol
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line
errors
256 = disable XON / XOFF
support
512 = disable hardware
FIFOs
1024 = enable RS-485
mode
4096 = enable command
prompt
(1)
Must power-cycle
instrument for these
options to take effect.
300,
RS-232 COM1 baud rate.
Enclose value in double
1200,
quotes (") when setting
2400,
from the RS-232
interface.
4800,
9600,
19200,
38400,
57600,
115200
Any character
RS-232 COM1 modem
in the allowed
initialization string. Sent
character set.
verbatim plus carriage
Up to 100
return to modem on
characters long.
power up or manually.
Enclose value in double
quotes (") when setting
from the RS-232
interface.
0 to 65535
RS-232 COM2 mode flags.
(Same settings as
RS232_MODE.)
045150110 Rev B.3