6.7.4.1. Analog Output Signal Type and Range Span Selection

To select an output signal type (DC Voltage or current) and level for one output channel,
activate the ANALOG I/O CONFIGURATION MENU (see Section 6.7.1) then press:
These keys set
the signal level
and type of the
selected channel

6.7.4.2. Analog Output Calibration Mode

The analog outputs can be calibrated automatically or manually. In its default mode, the
instrument is configured for automatic calibration of all channels. Manual calibration
should be used for the 0.1V range or in cases where the outputs must be closely
matched to the characteristics of the recording device. Outputs configured for automatic
calibration can be calibrated as a group or individually. Calibration of the analog outputs
needs to be carried out on first startup of the analyzer (performed in the factory as part
of the configuration process) or whenever re-calibration is required.
78
FROM ANALOG I/O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
CAL
SET>
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
DIAG AIO OUTPUT RANGE:
5V
0.1V
1V
5V
CURR
10V
DIAG AIO OUTPUT RANGE:
10V
0.1V
1V
5V
10V CURR
Model 100E Instruction Manual
EXIT
ENTR
EXIT
EXIT
EXIT
ENTR
EXIT
Pressing EXIT ignores the new setting and
ENTR EXIT
Press SET> to select the
analog output channel to be
configured. Press EDIT to
continue
Pressing ENTR records the new setting
and returns to the previous menu.
returns to the previous menu.
045150102 Rev XB1