Figure 3-1: Location Of Shipping Screws - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Getting Started
Verify that all of the optional hardware ordered with the unit has been installed. These
are checked on the paperwork (Form 04551) accompanying the analyzer.
Once you have determined that no shipping damage exists and the unit includes all
expected hardware options, remove three, RED colored shipping screws from the bottom
of the chassis, shown in Figure 3-1. There are:
Three locking down the sample chamber sensor housing assembly.,
Two locking down the internal pump visible from bottom of instrument).
Save these shipping screws and re-install them whenever the unit is shipped.
VENTILATION CLEARANCE: Whether the analyzer is set up on a bench or installed into
an instrument rack, be sure to leave sufficient ventilation clearance.
Back of the instrument
Sides of the instrument
Above and below the instrument.
26
Figure 3-1:
Location of Shipping Screws
Area
Model 100E Instruction Manual
Sensor Housing Shipping Screws
Remove from inside of instrument.
Pump Shipping Screws
Remove from outside, bottom of instrument.
NOTE
Minimum required clearance
10 cm / 4 inches
2.5 cm / 1 inch
2.5 cm / 1 inch
045150102 Rev XB1