TROUBLESHOOTING & REPAIR
are operating properly by checking the voltage test points on the relay board. Note
that the analyzer's DC power wiring is color-coded and these colors match the
color of the corresponding test points on the relay board.
Suspect a leak first! Customer service data indicate that half of all problems are
eventually traced to leaks in the pneumatic system of the analyzer (including the
external pump), the source of zero air or span gases or the sample gas delivery
system. Check for gas flow problems such as clogged or blocked internal/external
gas lines, damaged seals, punctured gas lines, a damaged pump diaphragm, etc.
Follow the procedures defined in Section 11.5 for confirming that the analyzer's
basic components are working (power supplies, CPU, relay board, keyboard, PMT
cooler, etc.). See Figure 3-8 for general layout of components and sub-assemblies
in the analyzer. See the wiring interconnect drawing and interconnect list, see
Appendix D.

11.1.1. Warning Messages

The most common and/or serious instrument failures will result in a warning message
displayed on the front panel. Table A-2 in Appendix A-3 contains a list of warning
messages, along with their meaning and recommended corrective action.
It should be noted that if more than two or three warning messages occur at the same
time, it is often an indication that some fundamental analyzer sub-system (power supply,
relay board, motherboard) has failed rather than an indication of the specific failures
referenced by the warnings. In this case, a combined-error analysis needs to be
performed.
The analyzer will alert the user that a warning is active by displaying the keypad labels
MSG and CLR on the front panel and a text message in the top center line of the display
as shown in this example:
The analyzer will also issue a message to the serial port and cause the red FAULT LED on
In WARNING mode, <TST TST>
keys replaced with TEST key.
Pressing TEST switches to
SAMPLE mode and hides warning
messages until new warning(s)
are activated.
If warning messages re-appear,
the cause needs to be found. Do
not repeatedly clear warnings
without corrective action.
198
SAMPLE
AZERO WARNING
< TST TST > CAL
SAMPLE
SYSTEM RESET
CAL
TEST
SAMPLE
RANGE=500 PPB
< TST TST > CAL
SAMPLE
SYSTEM RESET
< TST TST > CAL
Model 100E Instruction Manual
NOX =123.4
MSG CLR
SETUP
NOX = XXX.X
MSG
CLR
SETUP
NOX = XXX.X
CLR
SETUP
MSG
NOX = XXX.X
MSG
SETUP
CLR
045150102 Rev XB1
MSG indicates that one or more
warning message are active but
hidden. Pressing MSG cycles
through warnings
In SAMPLE mode, all warning
messages are hidden, but MSG
button appears
Press CLR to clear the current
warning message.
If more than one warning is
active, the next message will
take its place.
Once the last warning has been
cleared, the analyzer returns to
SAMPLE Mode.