Teledyne 100E Instruction Manual page 233

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
Setup Variable
DWELL_TIME
FILT_SIZE
FILT_ASIZE
FILT_DELTA
FILT_PCT
FILT_DELAY
FILT_ADAPT
DIL_FACTOR
USER_UNITS
LAMP_CAL
LAMP_GAIN
TEMPCO_GAIN
SLOPE_CONST
5
DARK_MODE
1, 3, 4
DARK_ENABLE
DARK_FREQ
045150110 Rev B.3
Numeric
Default Value
Units
Seconds
1,
4
0.2
Samples
240,
3
30
,
4
5
Samples
20,
3
6
PPM
0.02,
3
10
%
5,
2
10
Seconds
180
ON,
4
OFF
1
PPB,
3
PPM
3500
mV
0.9
0
8,
3
6.25
AUTO
ON,
4
OFF
Minutes
30,
3
720
APPENDIX A-2: Setup Variables For Serial
Value Range
0.1 to 10
1 to 480
1 to 100
0.001 to 0.1,
3
1 to 100
1 to 100
0 to 300
OFF, ON
0.1 to 1000
PPB,
PPM,
UGM,
MGM
1000 to 5000
0.5 to 1.5
0 to 2
0.1 to 10
OFF,
AUTO,
CMND
OFF, ON
0.1 to 1440
I/O, Revision B.3
Description
Dwell time before taking
each sample.
Moving average filter size.
Moving average filter size
in adaptive mode.
Absolute change to trigger
adaptive filter.
Percent change to trigger
adaptive filter.
Delay before leaving
adaptive filter mode.
ON enables adaptive
filter; OFF disables it.
Dilution factor if dilution
enabled with
FACTORY_OPT variable.
Concentration units for
user interface. Enclose
value in double quotes (")
when setting from the RS-
232 interface.
Last calibrated UV lamp
reading.
UV lamp compensation
attenuation factor.
Temperature coefficient
attenuation factor for
pressure readings.
Constant to make visible
slope close to 1.
Dark calibration mode.
OFF disables it; AUTO
does it periodically; CMND
does it only when
commanded. Enclose
value in double quotes (")
when setting from the RS-
232 interface.
ON enables PMT/UV dark
calibration; OFF disables
it.
Dark calibration period.
231