Optic Test - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual

6.7.5. Optic Test

The optic test function tests the response of the PMT sensor by turning on an LED located
in the cooling block of the PMT (Fig. 10-15). The analyzer uses the light emitted from the
LED to test its photo-electronic subsystem, including the PMT and the current to voltage
converter on the pre-amplifier board. To make sure that the analyzer measures only the
light coming from the LED, the analyzer should be supplied with zero air. The optic test
should produce a PMT signal of about 2000±1000 mV. To activate the electrical test
press the following key sequence.
SAMPLE*
RANGE = 500.000 PPB
< TST TST > CAL
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
SETUP X.X
ENTER DIAG PASS: 818
8
1
8
This is a coarse test for functionality and not an accurate calibration tool. The
resulting PMT signal can vary significantly over time and also changes with
045150102 Rev XB1
SO2 =X.XXX
SETUP
EXIT
MORE
EXIT
EXIT
ENTR
While the optic test is
activated, PMT should be
2000 mV ± 1000 mV
low-level calibration.
DIAG
SIGNAL I / O
JUMP
PREV NEXT
Press NEXT until...
DIAG
OPTIC TEST
PREV NEXT
RANGE = 500.000 PPB
DIAG OPTIC
<TST TST>
Press TST until...
DIAG ELEC
PMT = 2751 MV
<TST TST>
NOTE
Operating Instructions
ENTR
EXIT
EXIT
ENTR
SO2=X.XXX
EXIT
SO2=X.XXX
EXIT
85