Model 100E Instruction Manual
SIGNAL NAME
ST_ZERO_CAL
ST_SPAN_CAL
ST_DIAG_MODE
2,3,6,7
ST_LOW_SPAN_CAL
B status outputs, U27, J1018, pins 1–8 = bits 0–7, default I/O address 324 hex
ST_LAMP_ALARM
ST_DARK_CAL_ALARM
ST_FLOW_ALARM
ST_PRESS_ALARM
ST_TEMP_ALARM
ST_HVPS_ALARM
2
Front panel I
C keyboard, default I
MAINT_MODE
LANG2_SELECT
SAMPLE_LED
CAL_LED
FAULT_LED
AUDIBLE_BEEPER
Relay board digital output (PCF8575), default I
RELAY_WATCHDOG
RCELL_HEATER
045150110 Rev B.3
APPENDIX A-4: M100E Signal I/O Definitions,
BIT OR CHANNEL
NUMBER
3
4
5
6
7
0
1
2
3
4
5
6–7
2
C address 4E hex
5 (input)
6 (input)
8 (output)
9 (output)
10 (output)
14 (output)
2
C address 44 hex
0
1
2
DESCRIPTION
0 = in zero calibration
1 = not in zero
0 = in span calibration
1 = not in span
0 = in diagnostic mode
1 = not in diagnostic mode
0 = in low span calibration
1 = not in low span
Spare
0 = lamp intensity low
1 = lamp intensity OK
0 = dark cal. warning
1 = dark cal. OK
0 = any flow alarm
1 = all flows OK
0 = any pressure alarm
1 = all pressures OK
0 = any temperature alarm
1 = all temperatures OK
0 = HVPS alarm
1 = HVPS OK
Spare
0 = maintenance mode
1 = normal mode
0 = select second language
1 = select first language (English)
0 = sample LED on
1 = off
0 = cal. LED on
1 = off
0 = fault LED on
1 = off
0 = beeper on (for diagnostic testing only)
1 = off
Alternate between 0 and 1 at least every 5 seconds
to keep relay board active
0 = sample chamber heater on
1 = off
Spare
Revision B.3
243