3.2.3. Warning Messages

Because internal temperatures and other conditions may be outside of specified limits
during the analyzer's warm-up period, the software will suppress most warning
conditions for 60 minutes after power up.
If warning messages persist after 60 minutes, investigate their cause using the
troubleshooting guidelines in Section 11. The following table includes a brief description
of the various warning messages that may appear.
MESSAGE
ANALOG CAL WARNING
AZERO WRN XXX.X MV
BOX TEMP WARNING
CANNOT DYN SPAN
CANNOT DYN ZERO
CONFIG INITIALIZED
DARK CAL WARNING
DATA INITIALIZED
FRONT PANEL WARN
HVPS WARNING
IZS TEMP WARNING
PMT DET WARNING
PMT TEMP WARNING
RCELL PRESS WARN
RCELL TEMP WARNING
REAR BOARD NOT DET
RELAY BOARD WARN
SAMPLE FLOW WARN
SAMPLE PRESS WARN
SYSTEM RESET
UV LAMP WARNING
38
Table 3-7:
Possible Warning Messages at Start-Up
MEANING
The instruments A/D circuitry or one of its analog outputs is not
calibrated.
The reading taken during the auto-zero cycle is outside of specified limits.
The value XXX.X indicates the auto-zero reading at the time of the
warning.
The temperature inside the M200E chassis is outside the specified limits.
Remote span calibration failed while the dynamic span feature was set to
turned on
Remote zero calibration failed while the dynamic zero feature was set to
turned on
Configuration was reset to factory defaults or was erased.
Dark offset above limit specified
present in the sample chamber
iDAS data storage was erased.
Firmware is unable to communicate with the front panel.
High voltage power supply for the PMT is outside of specified limits.
On units with IZS options installed: The permeation tube temperature is
outside of specified limits.
output
PMT detector
PMT temperature is outside of specified limits.
Sample chamber pressure is outside of specified limits.
Sample chamber temperature is outside of specified limits.
The CPU is unable to communicate with the motherboard.
The firmware is unable to communicate with the relay board.
The flow rate of the sample gas is outside the specified limits.
Sample pressure outside of
The computer was rebooted.
The
intensity measured by the reference detector
UV lamp
too low or too high
Model 100E Instruction Manual
indicating that too much stray lightis
.
operational
outside of
operational parameters.
.
limits
reading
045150102 Rev XB1

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents