Operating Instructions
To disable a data channel, follow the instruction shown in section 6.10.2.2 then press:
Toggle key to turn
channel ON or OFF

6.10.2.11. HOLDOFF Feature

The iDAS HOLDOFF feature allows to prevent data collection during calibrations and
during the DAS_HOLDOFF period enabled and specified in the VARS (Section 6.6). To
enable or disable the HOLDOFF, follow the instruction shown in section 6.10.2.2 then
press:
Toggle key to turn
HOLDOFF ON or OFF
116
FROM THE DAS PASSWORD MENU
Edit Data Channel Menu
SETUP X.X
0) CONC:
ATIMER,
PREV NEXT
INS
DEL
SETUP X.X
NAME:CONC
<SET
EDIT
PRINT
SET>
Press SET> key until...
SETUP X.X
CHANNEL ENABLE:ON
<SET
SET>
PRINT
EDIT
SETUP X.X
CHANNEL ENABLE:ON
OFF
FROM THE DAS PASSWORD MENU
Edit Data Channel Menu
SETUP X.X
0) CONC:
ATIMER,
PREV NEXT
INS
DEL
EDIT
SETUP X.X
NAME:CONC
<SET
EDIT
PRINT
SET>
Press SET> key until...
SETUP X.X
CAL HOLD OFF:ON
SET>
PRINT
EDIT
SETUP X.X
CAL HOLD OFF:ON
ON
Model 100E Instruction Manual
2,
900
PRNT
EDIT
EXIT
EXIT
EXIT
ENTR
EXIT
2,
900
PRNT
EXIT
EXIT
EXIT
ENTR
EXIT
Exits to the main
Data Acquisition
menu
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
Exits to the main
Data Acquisition
menu
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
045150102 Rev XB1