6.7.10. Test Channel Output

When activated, output channel A4 can be used to report one of the test functions
viewable from the SAMPLE mode display. To activate the A4 channel and select a test
function, follow this key sequence:
SAMPLE*
RANGE = 500.000 PPB
< TST TST > CAL
PRIMARY SETUP MENU
SETUP X.X
CFG DAS RNGE PASS CLK
SECONDARY SETUP MENU
SETUP X.X
COMM VARS
DIAG
DIAG
SIGNAL I / O
NEXT
DIAG
ANALOG OUTPUT
PREV
NEXT

Table 6-13: Test Parameters Available for Analog Output A4

Once a TEST function is selected, the instrument begins to report a signal on the A4
output and adds TEST= to the list of test functions viewable on the display (just before
the TIME display).
045150102 Rev XB1
SO2 =X.XXX
SETUP
EXIT returns
to the main
EXIT
MORE
SAMPLE
display
EXIT
ENTR EXIT
ENTR EXIT
TEST CHANNEL
NONE
PMT READING
UV READING
SAMPLE PRESSURE
SAMPLE FLOW
RCELL TEMP
CHASSIS TEMP
IZS TEMP
PMT TEMP
CHASSIS TEMP
HVPS VOLTAGE
1
This refers to the voltage range of the parameter and
not the output signal of the test channel.
Continue to press NEXT until ...
DIAG
PREV
NEXT
DIAG TCHN
NEXT
DIAG TCHN
TEST CHANNEL: PMT DETECTOR
PREV
NEXT
Press PREV or NEXT
to move through the
list of available
parameters
(Table 6-9)
TEST PARAMETER RANGE
Test channel is turned off
0-5000 mV
0-5000 mV
0-40
in-Hg-A
0-1000 cm³/min
0-70
° C
0-70
° C
0-70
° C
0-50
° C
0-70
° C
0-5000 V
Operating Instructions
TEST CHAN OUTPUT
EXIT
ENTR
TEST CHANNEL: NONE
ENTR
EXIT
ENTR
EXIT
Press ENTR to
Press EXIT to
select the displayed
return to the
parameter and to
DIAG menu
activate the test
channel.