Viewing Idas Data And Settings - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
These default Data Channels can be used as they are, or they can be customized from
the front panel to fit a specific application. They can also be deleted to make room for
custom user-programmed Data Channels.
Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text
file can either be loaded into APICOM and then modified and uploaded to the instrument
or can be copied and pasted into a terminal program to be sent to the analyzer.
Sending an iDAS configuration to the analyzer through its COM ports will
replace the existing configuration and will delete all stored data. Back up any
existing data and the iDAS configuration before uploading new settings.

6.10.2.1. Viewing iDAS Data and Settings

iDAS data and settings can be viewed on the front panel through the following keystroke
sequence.
EXIT will return to the
main SAMPLE Display.
045150102 Rev XB1
SAMPLE*
RANGE = 500.000 PPB
< TST TST > CAL
PRIMARY SETUP MENU
SETUP X.X
CFG
RNGE PASS CLK MORE
DAS
SETUP X.X
DATA ACQUISITION
EDIT
VIEW
SETUP X.X
: DATA AVAILABLE
CONC
NEXT
VIEW
SETUP X.X
ENGDAT: DATA AVAILABLE
PREV
NEXT
VIEW
SETUP X.X
CALDAT: DATA AVAILABLE
PREV
VIEW
NOTE
SO2 =X.XXX
SETUP
EXIT
EXIT
Keys only appear as needed
EXIT
SETUP X.X
00:00:00
NXCNC1=0.0 PPM
<PRM
NEXT NX10
PV10 PREV
EXIT
SETUP X.X
00:00:00 SMPFLW=000.0 cc / m
<PRM
EXIT
SETUP X.X
00:00:00
NXSLP1=0.000
<PRM
PV10 PREV
Operating Instructions
VIEW KEYPAD FUNCTIONS
KEY
FUNCTION
<PRM
Moves to the next Parameter
PRM>
Moves to the previous
Parameter
NX10
Moves the view forward 10
data points/channels
NEXT
Moves to the next data
point/channel
PREV
Moves to the previous data
point/channel
PV10
Moves the view back 10 data
points/channels
EXIT
PRM>
EXIT
PRM>
EXIT
PRM>
107