3.5.2. Warm-Up

The M200E requires about 30 minutes warm-up time before reliable NO
measurements can be taken. During that time, various portions of the instrument's front
panel will behave as follows. See Figure 3-4 for locations.
Table 3-5:
Name
Color
Concentra-
N/A
tion Field
Mode Field
N/A
STATUS LEDs
Sample
Green
Cal
Yellow
Fault
Red

3.5.3. Warning Messages

During warm-up, internal temperatures and other parameters may be outside of specified
limits. The software will suppress most warning conditions for 30 minutes after power up.
044100102 Rev A
:
:
:
:
M200E NOX ANALYZER
SOFTWARE REVISION C.4
SYSTEM RESET
NOX=X.X
CAL
CLR SETUP
Press CLR to clear initial warning messages.
(see Section 3.2.3)
Front Panel Display During System Warm-Up
Behavior
Switches between
NO
, NO and NO
X
2
Displays blinking
"SAMPLE"
On
Off
Blinking
System waits 3 seconds then
3
automatically begins its initialization
routine. No action required.
.
1
System is checking the format of
the instrument's flash memory chip.
If at this point,
1
**FLASH FORMAT INVALID**
appears, contact T–API customer
service
1
The instrument is loading
configuration and calibration data
from the flash chip
The instrument is loading the
analyzer firmware.
The revision level of the firmware
installed in your analyzer is briefly
displayed
Firmware
fully booted
Significance
This is normal operation.
Instrument is in sample mode but is still in the
process of warming up (hold-off period is
active).
Unit is operating in sample mode, front panel
display is continuously updated.
The instrument's calibration is not enabled.
The analyzer is warming up and out of
specification for a fault-free reading.
Getting Started
, NO and NO
X
2
21