Press the CLR key to clear the SYSTEM RESET warning. The display should look
like this:
SAMPLE
<TST TST>
NOTE: The word SAMPLE in the upper right corner of the display may flash on and
off for several minutes as the unit warms up. This is normal.

3.2.2. Warm Up

The M300E requires about 30 minutes for all internal components to come up to
temperature before reliable CO measurements can be taken.

3.2.3. Warning Messages

Because internal temperatures and other conditions may be outside be specified
limits during the analyzers warm-up period, the software will suppress most
warning conditions for 30 minutes after power up.
If warning messages persist after the 30 minutes warm up period is over,
investigate their cause using the troubleshooting guidelines in Section 11 of this
manual. The following table includes a brief description of the various warning
messages that may appear.
Message
SYSTEM RESET
SOURCE
SYNC
BENCH TEMP
WHEEL TEMP
BOX TEMP
SAMPLE TEMP
SAMPLE FLOW
SAMPLE PRES
P/N 04288 REV: A
RANGE=50.0 PPM
CAL
The computer has rebooted.
The IR source may be faulty.
The instrument is not properly tracking the rotation of the Gas Filter
Correlation wheel.
Optical bench temperature is outside the specified limits.
The Gas Filter Correlation wheel temperature is outside the specified
limits.
The temperature inside the M300E chassis is outside the specified
limits.
The temperature of the sample gas is outside the specified limits.
The flow rate of the sample gas is outside the specified limits.
The pressure of the sample gas is outside the specified limits.
CO=00.00
Meaning
Getting Started
SETUP
33