Specifications, Approvals and Warranty
2.2. EPA Equivalency Designation
The Model 200E Analyzer is designated as Reference Method Number RFNA-1194-099
(same designation as model M200A) as per 40 CFR Part 53 when operated under the
following conditions:
•
Range: Any full-scale range between 0-0.05 and 0-1.0 ppm (parts per million).
•
Ambient temperature range of 5 to 40
•
Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
•
Equipped with PTFE filter element in the internal filter assembly.
•
Equipped with ozone supply air filter
External vacuum pump capable of 10 in-Hg-A at 2 standard liters per minute (slpm) or
better.
Software settings:
Dynamic span
Dynamic zero
Cal-on-NO
Dilution factor
Temp/Pressure compensation
AutoCal
Independent range
Auto-range
Converter efficiency
Under the designation, the analyzer may be operated with or without the following optional
equipment:
•
Rack mount with or without slides.
•
Rack mount for external pump.
•
Zero/span valve options.
•
Nafion-type permeation sample gas conditioner
•
Internal zero/span (IZS) option with:
•
NO
permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.
2
•
NO
permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified.
2
Under the designation, the IZS option cannot be used as the source of calibration.
•
4-20mA isolated analog outputs.
•
Status outputs.
•
Control inputs.
•
RS-232 output.
•
Ethernet communications option
6
2
Model 200E Instruction Manual
o
C.
OFF
OFF
OFF
OFF
ON
ON or OFF
ON or OFF
ON or OFF
Acceptable values of 0.96 to 1.02
044100102 Rev A