Specifications, Agency Approvals, and
Warranty

2.2. EPA Equivalency Designation

Advanced Pollution Instrumentation, Inc., Model 300E Carbon monoxide Analyzer is
designated as Reference Method Number EQOA-0992-087 as defined in 40 CFR Part
53, when operated under the following conditions:
1.
Range: Any range from 10 ppm to 50 ppm
2.
Ambient temperature range of 10 to 40°C
3.
Line voltage range of 90 – 127 and 200 – 230 VAC, 50/60 Hz
4.
With 5-micron TFE filter element installed in the internal filter assembly
5.
Sample flow of 800 ± 80 cc/min at sea level
6.
Internal sample pump
7.
Software settings:
Dilution factor
AutoCal
Dynamic Zero
Dynamic Span
Dual range
Auto range
Temp/Pres compensation
Under the designation, the Analyzer may be operated with or without the following
options:
1.
Rack mount with slides
2.
Rack mount without slides, ears only
3.
Option 52 Sample/Cal valves
4.
Option 53 Sample/Cal valves with internal zero air
5.
Option 50 Sample/Cal valves with span shutoff & flow control
6.
Option 51 Sample/Cal valves with span shutoff & flow control & internal
zero air
7.
4-20mA, isolated output
18
Model 300E Carbon Monoxide Analyzer
1.0
ON or OFF
ON or OFF
OFF
ON or OFF
ON or OFF
ON
Instruction Manual
P/N 04288 REV: A