Sections 7.2 and 7.3 describe procedures for performing manual Zero and Span Calibrations on
your analyzer.
To perform a CHECK of the instrument's Zero or Span Calibration follow the same steps,
however, DO NOT press the ENTR key at the end of each operation.
Pressing the ENTR key resets the stored values for OFFSET and SLOPE and alters the
instruments Calibration.
When performing a CHECK ZERO, press the EXIT key to end the procedure.
7.2. Manual Calibration without Zero/Span
Valves
This is the basic method for manually calibrating the Model 300E CO Analyzer
without functioning Zero/Span Valve options. It is identical to the method
described in the GETTING STARTED section of this manual and is repeated her for
you convenience.
STEP ONE: Connect the Sources of Zero Air and Span Gas as shown below.

Figure 7-1: Pneumatic Connections for Manual Calibration w/o Z/S Valves

P/N 04288 REV: A
/SPAN CALIBRATION VS. ZERO/SPAN CHECK
Calibrated
CO Gas
MODEL 701
Zero Air Generator
Calibrated
CO Gas
Valve
MODEL 701
Zero Air Generator
No Valve Options Installed
MODEL 700
Gas Dilution
Calibrator
VENT
Sample
MODEL
Pressure Span
Vent Span
IZS
Exhaust
OR
Needle Valve to
control Flow
Sample
Pressure Span
Vent Span
VENT
IZS
Exhaust
CALIBRATION PROCEDURES
Source of
SAMPLE Gas
Removed
during
Calibration
300E
Source of
SAMPLE Gas
Removed
during
Calibration
MODEL
300E