P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-5-4
T-TRx Box Replacement Procedure
CAUTION
8-5-4-1
Tools
Appropriate Phillips screwdriver.
FRU Part #
Table 9-12
Refer to
8-5-4-2
Time Required
15 minutes
8-5-4-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-5-4-4
T-TRx Box Removal Procedure
1.) Remove all accessories.
•
2.) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, and Right
Side eTower.
•
•
•
•
3.) Open the two fastening latches on the left side of the FE door assembly and loosen the captive
screw. Then, carefully open the door.
4.) Disconnect the MPB TO T-FEPS Cable.
5.) Disconnect the fan cable.
6
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
.
on page 9-12
Accessories - Replacement Procedures
RS Probe Cover Installation Procedure
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL
8-73