Functional Checks - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
10-5-2

Functional Checks

NOTE:
Refer also to
functional checks described in this section.
The functional checks take approximately 60 minutes to perform. Refer to the Venue™ User Manual
whenever necessary.
10-5-2-1
System Checks
Step
1
B-Mode
2
CF-Mode
3
Doppler Modes
4
M-Mode
5
Basic Measurements
6
Applicable Software Options
7
Monitor
8
Peripherals
6
EVISION
Chapter 4 - General Procedures and Functional
Table 10-5
System Functional Checks
Item (or Mode)
Verify basic B-Mode (2D) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic CF-Mode (Color Flow Mode) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic Doppler operation (PW and CW if available).
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic M-Mode operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Check Distance and Tissue Depth Measurement.
Verify the basic operation of all optional modes such as Contrast. Check the basic
Ultrasound system controls that affect each option's operation.
Verify basic monitor display functions.
See:
Chapter 10 - Care and Maintenance
Peripheral/Option Checks
on page 10 - 8.
V
™ S
ENUE
Checks, for additional details about the
Description
M
ERVICE
ANUAL
10-7