Required Facility Needs - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
2-3-2

Required Facility Needs

The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in section
Requirements
A dedicated "hospital-grade" single branch power outlet of adequate amperage (see
page
unit's proposed location. Refer to the
A door opening of at least 76 cm (30 in) wide.
The proposed location for the unit is at least 0.5 m (1.5 ft) from the walls, to enable cooling.
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the Venue™ system to connect cables.
Power outlets for other medical equipment.
Power outlets for test equipment within 1 m (3.2 ft) of the ultrasound unit.
Clean and protected space to store probes (in their cases or on a rack).
Material to safely clean probes (done using a plastic container, never metal).
In the case of a network option:
NOTE:
All relevant preliminary network outlets installations at the prepared site must be performed by
authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform
servicing of the equipment.
2-8
6
EVISION
(see below).
2-3.) that meets all local and national codes and is located less than 2.5 m (8.2 ft) from the
An active network outlet in the vicinity of the ultrasound unit.
A network cable of appropriate length (regular Pin-to-Pin network cable).
An IT administrator who will assist in configuring the unit to work with your local network. A fixed
IP address may be required when using DICOM. Refer to the form provided in
page 2-13 for network details that are required.
Section 2-3 - Facility Needs
2-3-3 - Networking Pre-Installation
Electrical Requirements
V
™ S
ENUE
ERVICE
Table 2-3 on
section on page 2-3.
Figure 2-3
M
ANUAL
on

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Venue r2H45051vnH45281vn

Table of Contents