Replacement of the Probe Mux Box

8-19-13-1

Manpower

One person, 10 minutes
8-19-13-2

Tools

Phillips screwdriver 1
Socket wrench 4
8-19-13-3

Preparations

1.) Power Off/Shutdown the system as described in
2.) Remove the Probe Holder as described in
8-19-13-4

Probe Mux Box - Removal Procedure

1.) Disconnect the Probe connector.
2.) Hold the Probe box to the Dock Cart.
10
EVISION
Figure 8-77 disconnect the Probe connector
Chapter 8 - Replacement Procedures
Section 3-6-3 on page
Section 8-19-11 on page
V
i / V
OLUSON
OLUSON
S
M
ERVICE
3-72.
8-55.
e
ANUAL
8-57