P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-7-3
BE to Cockpit Cable Replacement Procedure
8-7-3-1
Tools
•
Appropriate Phillips screwdriver
•
Flat screwdriver
•
2.5 mm Allen key.
FRU Part #
Table 9-14
Refer to
8-7-3-2
Time Required
15 min
8-7-3-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-3-4
BE to Cockpit Cable Removal Procedure
1.) Remove all accessories.
•
2.) Remove the following covers: RS Probe, Lower Front eTower, Left side eTower, Right Side eTower,
Riser Thermal, and MPB Door cover,
•
•
•
•
•
•
3.) Disconnect the BE TO COCKPIT CABLE from the monitor:
•
•
Figure 8-116 Cockpit (Monitor) Cable and Cockpit Ground Cables Removal
6
EVISION
.
on page 9-14
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Riser Thermal Cover Removal Procedure
MPB Front Metal Door Removal Procedure
Disconnect the cable connector (use flat-head screwdriver)
Release Ground cable (use Phillips screwdriver).
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL
8-123