Cockpit (Monitor) Replacement Procedure - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-4-5

Cockpit (Monitor) Replacement Procedure

8-4-5-1
Tools
Appropriate Phillips screwdriver
Flat-head screwdriver
FRU Part #
Refer to
8-4-5-2
Time Required
15 min
8-4-5-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-5-4
Cockpit (Monitor) Removal Procedure
1.) On the back of the cockpit, release four screws and remove the transparent Wi-Fi dongle cover.
2.) Remove four Phillips screws securing the cockpit (monitor) to the arm.
8-36
6
EVISION
Table 9-11
.
on page 9-11
Disconnect the Wi-Fi dongle.
Figure 8-22 Wi-Fi Cover and Dongle Removal
Section 8-4 - System Modules- Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL

Hide quick links: