Plastic Cable Guide For Cfe Replacement Procedure - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-6-7

Plastic Cable Guide for CFE Replacement Procedure

8-6-7-1
Tools
Appropriate Phillips screwdriver
FRU Part #
Table 9-5
Refer to
8-6-7-2
Time Required
15 min
8-6-7-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-7-4
Plastic Cable Guide for CFE Removal Procedure
1) Remove all accessories.
2) Remove the following covers: RS Probe cover, Lower Front eTower, Left side eTower, Right Side
eTower, Riser Thermal Cover
3.) Open the two fastening latches on the left side of the FE door assembly, loosen the captive screw
and carefully open the door. See
4.) Secure the door by clipping the door with door support-
5.) Remove the BIB cable from the Plastic Cable Guide for CFE as follows:
6
EVISION
.
on page 9-4
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Riser Thermal Cover Removal Procedure
Release the two screws.
Chapter 8 - Replacement Procedures
Figure
8-19.
Figure
V
™ S
ENUE
Power Shut Down
.
8-19.
M
ERVICE
ANUAL
8-97