Psu Module Replacement Procedure - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-4-6

PSU Module Replacement Procedure

8-4-6-1
Tools
Appropriate Phillips screwdriver.
FRU Part #
Refer to
8-4-6-2
Time Required
15 min
8-4-6-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-6-4
PSU Module Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower cover, Right Side eTower
Cover and the Left Side eTower Cover:.
3.) Remove the MPB door.
4.) Remove the two battery packs
8-40
6
EVISION
Table 9-10
.
on page 9-10
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
MPB Door Cover Removal Procedure
Battery Module Removal Procedure
Section 8-4 - System Modules- Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL