Full Front End Replacement
8-4-3-1
Tools
Appropriate Phillips screwdriver
FRU Part #
Table 9-12
Refer to
8-4-3-2
Time Required
5 min
8-4-3-3
Preparation
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-3-4
Full Front End Removal
1) Remove all accessories.
•
2) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, and Right
Side eTower.
•
•
•
•
3.) Open the two fastening latches on the left side of the FE door assembly and loosen the captive
screw. Then, carefully open the door.
6
EVISION
.
on page 9-12
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Figure 8-19 Opening FE Door
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL
8-31