P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-7-4
PSU to MPB DC Docking Cable Replacement Procedure
8-7-4-1
Tools
•
Appropriate Phillips screwdriver
•
Flat screwdriver.
FRU Part #
Table 9-14
Refer to
8-7-4-2
Time Required
15 min
8-7-4-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-4-4
PSU to MPB DC Docking Cable Removal Procedure
1.) Remove all accessories.
•
2.) Remove the MPB module.
•
3.) Remove the fan.
•
4.) Disconnect J8 to lower PSU connectors (see connectors' diagram below):
6
EVISION
.
on page 9-14
Accessories - Replacement Procedures
MPB Module Removal Procedure
MPB Blower (Fan) Removal Procedure
Figure 8-134 Connectors Diagram
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL
8-133