P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-7-7
PCIe Cable Replacement Procedure
8-7-7-1
Tools
Appropriate Phillips screwdriver.
FRU Part #
Refer to
8-7-7-2
Time Required
10 min
8-7-7-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-7-4
PCIe Cable Removal Procedure
1.) Remove all accessories.
•
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower.
•
•
•
3.) Remove the BE module.
•
4.) Disconnect one side of the PCIe cable from the BIB board and the other side from the BE module.
5.) Remove the PCIe cable.
8-142
6
EVISION
Table 9-14
.
on page 9-14
Accessories - Replacement Procedures
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
BE Module Removal Procedure
PCIe cable location
Figure 8-144 Disconnecting PCIe Cable
Section 8-6 - Mechanical Parts- Replacement Procedures