GE Venue R1 Service Manual page 98

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
3-5-5-2
Connect a Probe
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe.
CAUTION
CAUTION
CAUTION
There are 4 probe connectors on the rear panel of the system, numbered 1 through 4. Each connector
is comprised of a probe-socket (6) and a locking latch (5).
1) Before connecting the probe:
a.) Do a visual check of the probe pins and system sockets.
b.) Remove any dust or foam rests from the probe pins.
c.) Verify the probe and the probe cable for any visual damage.
6
EVISION
Do not allow the probe head to hang freely. Excessive impact
to the probe will result in irreparable damage.
To prevent probe connector pins damage, or PCB board
damage, do not use excessive force when connecting the
probes.
Keep the probe cables away from the wheels.
Do not bend the probe cables.
Do not cross cables between probes.
Figure 3-26
Chapter 3 - System Setup
3
Venue™ Probe Connectors
V
™ S
M
ENUE
ERVICE
ANUAL
3-25