MPB to BIB Control Cable Replacement Procedure

8-7-8-1
Tools
Appropriate Phillips screwdriver.
FRU Part #
Refer to
8-7-8-2
Time Required
10 min
8-7-8-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-8-4
MPB to BIB Control Cable Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower.
3) Remove MPB Front Metal Door.
4.) Remove the MPB module.
5.) Disconnect the cable from the BIB board.
8-144
6
EVISION
Table 9-14
.
on page 9-14
Accessories - Replacement Procedures
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
MPB Front Metal Door Removal Procedure
MPB Module Removal Procedure
Figure 8-145 Disconnecting MPB to BIB Cable