Mpb To Be Pwr Cable Replacement Procedure - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-7-5

MPB to BE PWR Cable Replacement Procedure

8-7-5-1
Tools
Appropriate Phillips screwdriver
Flat screwdriver.
FRU Part #
Refer to
8-7-5-2
Time Required
15 min
8-7-5-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-5-4
MPB to BE PWR Cable Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe cover, Lower Front eTower, Side Covers, MPB Power Door
3.) Remove the MPB Module.
8-136
6
EVISION
Table 9-14
.
on page 9-14
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
MPB Door Cover Removal Procedure
MPB Module Removal Procedure
Section 8-6 - Mechanical Parts- Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL