Riser Cover Replacement Procedure; Irection Revision Enue Ervice Anual - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-3-11

Riser Cover Replacement Procedure

8-3-11-1
Tools
Appropriate Phillips screwdriver
FRU Part #
Refer to
8-3-11-2
Time Required
5 min
8-3-11-3
Preparations
Shut down the Venue™ ultrasound unit, as described in
Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-3-11-4
Riser Cover Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, RS Probe Cover. Right Side and Left Side
eTower covers, Upper Front eTower, and Riser Thermal Cover.
3.) Remove cockpit (monitor).
4.) Remove Scanner Arm.
5.) Loosen and remove four screws securing the riser cover to the riser (access screws from
underneath).
8-20
6
EVISION
Table 9-9
.
on page 9-8
Accessories - Replacement Procedures
Lower Front eTower Cover Removal Procedure
RS Probe Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Upper eTower Front Cover Removal Procedure
Riser Thermal Cover Removal Procedure
Cockpit (Monitor) Removal Procedure
Arm Removal Procedure
Section 8-3 - Covers - Replacement Procedures
V
ENUE
Power Shut Down
™ S
M
ERVICE
ANUAL