Cooling System

5-11-1

General Information

The Venue™ system has 3 fans for system cooling:
Back End Fan - located on the Back End module.
T-CFE Fan - located on the top of the Front End module.
MPB Blower (Fan) - located on the on the rear side of the system behind the MPB module.
All fans are controlled and have variable speed.
The cooling requirement for the Venue™ ultrasound scanner with monitor and on board peripherals, is
up to 2000 BTU/h. This figure does not include cooling needed for lights, people, or other equipment in
the room.
NOTE:
Each person in the room places an additional 300 BTU/h demand on the cooling system.